beautypg.com

D.1 wiring – Campbell Scientific 43347 RTD Temperature Probe and 43502 Aspirated Radiation Shield User Manual

Page 36

background image

Appendix D. Measure Two 43347-IX Probes Using One Current Excitation Channel

D.1 Wiring

Wiring for two 43347-IX probes is shown in FIGURE D-1.

Ground CLEAR

Current Excite/+ RTD RED

Sense Signal

WHITE

Sense Signal Ref

GREEN

BLACK

EARTH GND

+ RTD

+ SENSE

- SENSE

- RTD

1000 OHM
RTD

Wire Label

43347

Terminals

R

s

Ground CLEAR

RED

Sense Signal

WHITE

Sense Signal Ref

GREEN

EARTH GND

+ RTD

+ SENSE

- SENSE

1000 OHM
RTD

43347

Terminals

R

s

#1

#2

Current Return/- RTD BLACK

- RTD

FIGURE D-1. Schematic for Two 43347-IX Temperature Probes

D.2 Example Program for two Calibrated 43347-IX

Probes

This section includes an example CR3000 program that measures two
calibrated 43347-IX probes. A CR5000 is programmed similarly. Wiring for
the example program is shown in TABLE D-1.

D-2