Appendix C. Sensor Maintenance

C.1 Reed Switch Replacement Procedure

To verify parts and locations, refer to the cable and parts diagrams (FIGURE

C-2 and FIGURE C-3) and the parts list (TABLE C-1).

A. Remove sensor from mounting arm and disconnect cable.

B. Remove the cup assembly.

C. Remove the three Philips screws at the top of the sensor and lift out the

bearing mount assembly.

D. Unsolder the leads of the reed switch and remove the switch from the two

mounting terminals, see the parts diagram.

E. Solder the new switch onto the sides of the switch mount terminals (form

a loop in the relay leads to obtain proper lead length --

DO NOT CUT

THE RELAY LEADS.) Measure the distance between the bottom of the

rotating magnet and the top of the switch envelope, as shown in FIGURE

C-1. The spacing should measure between 0.01 and 0.02 inches.

F. Spin the shaft to verify switch operation by listening for a faint sound of

the switch closure. If the switch cannot be heard, move the switch slightly

closer to the magnet assembly.

G. Reassemble sensor.

FIGURE C-1. Reed switch assembly

C.2 Bearing Replacement Procedure

The bearings used in the 014A sensor are special stainless steel ball bearings

with a protective shield. Bearings are lubricated and sealed.

DO NOT

LUBRICATE BEARINGS AS THE LUBRICATION WILL ATTRACT

DUST AND INHIBIT BEARING OPERATION.

A. Follow steps A, B, and C in reed switch replacement procedures.

B. Loosen set screws in magnet assembly, lift shaft and collar up and out of

bearing mount. Be sure to retain lower spacer.

C-1