beautypg.com

Ashcroft 2084 Precision Digital Test Gauges User Manual

Page 12

background image

– 12 –

To access the factory default password:

Step 1: Press the CONFIG key until the word

CALIBRAT appears.

Step 2: Press the ENTER key.

Step 3: The letters/asterisks… PW***** appear.

Step 4: Press the CONFIG key. An Ø appears in

the first position.

Step 5: Press the ENTER key once.

Step 6: Press the CONFIG key until Ø appears. Ø

will appear in the second position.

Step 7: Press ENTER.

Step 8: Use this format until all the asterisks are

replaced with Ø.

There now should be a total of five Ø’s on the
keyboard display. The zero in the fifth position
should be blinking.

Step 9: Press the ENTER key. You are now

prompted to SET PW (or set password).

Step 10: Press the ENTER key.

Step 11: Decide on a five number user password,

then follow the procedure above insert-
ing a number in the flashing display until
all five numbers are inserted.

Step 12: A SAVE prompt will then appear. If the

selected user password is acceptable,
press ENTER. If the selected user pass-
word is not acceptable press ZERO
CLEAR to refigure the user password.

After the password is configured, the default
factory password will be replaced with the user
password. Once configured, the factory password
is no longer accessible.

If an incorrect password is entered, the system
will display WRONG. Press the CONFIG key to
reenter the correct password.

Step 13: Press ENTER again to begin calibration.

Note: Calibration feature allows recalibration of
zero and span.

*Indicates default.

KEYPAD FUNCTIONS

This manual is related to the following products: