Ashcroft 2084 Precision Digital Test Gauges User Manual
Page 11

– 11 –
DAMPING: with six different options, this mode
allows for taking process pressure readings and
averaging them. This option is particularly useful
when there is pulsation in the process. The
options are NONE*, AVG 2, 4, 6 or 8.
Step 1: Press the CONFIG key until the word
DAMPING appears.
Step 2: Press the ENTER key.
Step 3: Press the CONFIG key to select one of the
available DAMPING options.
Step 4: Press the ENTER key to finalize your
DAMPING option.
CONTRAST: this mode allows for BACKLITE
contrast level. Seven options are available,
1, 2, 3, 4*, 5, 6 and 7.
Step 1: Press the CONFIG key until the word
CONTRAST appears.
Step 2: Press the ENTER key.
Step 3: Press the CONFIG key to select one of the
available CONTRAST options.
Step 4: Press the ENTER key to finalize your
CONTRAST selection.
Note: setting high contrast levels will decrease
battery life.
CALIBRAT.:
Gauge Calibration: Both zero and span adjust-
ments are available. This gauge has been config-
ured with a default password of ШШШШШ. This fac-
tory password does not allow access to calibra-
tion. To access the calibration mode, it is neces-
sary to configure a user password. Once the user
password is configured, it will become the default
password that allows access to gauge calibration.
*Indicates default.
KEYPAD FUNCTIONS