5

Installation
IMPORTANT

Be sure to read & understand all of the Instructions

before beginning.

Unpacking

Carefully remove the contents of the shipping

carton and check each item against the packing

list before destroying any packing materials.

Storage

Level gauge probes should be stored in their original

shipping containers until ready for installation.

Damage that occurs in storage is not covered under

manufacturer warranty.

Mounting Conditions
CAUTION

• When installing probes, do not bend rigid
probes, permanent damage may result.
• Longer rigid probes need to be supported at
both ends while handling.
• Probes are sealed at the factory and
contain no user serviceable components.
• Do not attempt to open probe or weld the
tube.

• Level gauge probes are designed for industrial

applications, but should be mounted in a

location as free as possible from vibration,

corrosive atmospheres, or any possibility of

mechanical damage.

• Place the level gauge in a reasonably accessible

location, ambient temperature should be

between -40°F and 158°F (-40°C to 70°C).

• Mount the Level gauge probe perpendicular with

gravity.

• Float should have free movement along probe.

• Float retention clip should be in place at base of

probe.

Mounting Considerations

Mounting considerations may vary (Flanges,

Compression Fitting, etc.) depending on the

application. For underground tanks, the probe is

generally mounted in the riser, resting on the bottom

of the tank. Spacers are used to hold the sensor in

the center of the riser. While most underground tanks

are horizontal and fairly standard in design, above

ground tanks vary considerably. The requirements

for mounting these probes are fairly simple. Since the

probe requires a float to provide level position, there is

a minimum size required for insertion of the float into

the tank.

It is recommended that a minimum of a 2” NPT pipe

opening be used.