Chapter 6 ____________________________________________________ Calibration and adjustment

VAISALA ________________________________________________________________________ 61

Relative Humidity Calibration after

Sensor Change

Syntax: FCRH

The transmitter asks and measures relative humidity and calculates the

calibration coefficients. This two-point adjustment needs to be performed

after a sensor change. Follow the more detailed calibration instructions

on page 56, but instead of the CRH command, use the FCRH command.

>FCRH

RH

:

1.82 1. ref

?

0

Press any key when ready...

RH

: 74.22

2. ref

? 75

OK

>
OK

indicates that the calibration has succeeded.

Set Calibration Information Text

Syntax: CTEXT

1.

Remove the transmitter unit from the mounting (see page 18) and

press the adjustment button once (see Figure 15 on page 56).

2.

Issue the LI command and enter value 0 for offset and value 1 for

gain.

3.

Issue the CTEXT command and enter calibration information text.

Set Calibration Date

Syntax: CTEXT

1.

Remove the transmitter unit from the mounting plate (see section

Mounting the Transmitter/Removing the Transmitter Unit on page

18) and press the adjustment button once (see Figure 15 on page

56).

2.

Issue the LI command and enter value 0 for offset and value 1 for

gain.

3.

Issue the CDATE command and set the calibration date.

Example:

CDATE 2001 12 11

Calibration : 2001-12-11

>