Scale analog outputs, Test analog outputs – Vaisala MMT310 User Manual
Page 46

User's Guide _______________________________________________________________________
46 ___________________________________________________________________ M210474EN-C
Scale Analog Outputs
Syntax: ASCL
Example:
>ascl
Ch1 T lo : -40.00 'C ? -20
Ch1 T hi : 60.00 'C ? 40
Ch2 ppm lo : 0.00 ppm ? 0
Ch2 ppm hi : 5000.00 ppm ? 3000
>ascl
Ch1 T lo : -20.00 'C ?
Ch1 T hi : 40.00 'C ?
Ch2 ppm lo : 0.00 ppm ?
Ch2 ppm hi : 3000.00 ppm ?
The ppm calculation is only used for transformer oils.
Test Analog Outputs
Syntax: ITEST aa.aaa bb.bbb
The operation of the analog outputs is tested by forcing the outputs to
given values. The values in the analog outputs can then be measured with
a current/voltage meter.
Where
aa.aaa
= Current/voltage value to be set for channel 1
bb.bbb
= Current/voltage value to be set for channel 2
Example:
>itest 8.30 6.40
8.30000mA 2c30 6.40000mA 2278
>
The ITEST command output includes internal debugging values shown
in hexadecimal format (
2c30
and
2278
in the example above).
The set values remain valid until you issue the ITEST command without
readings or reset the transmitter.