Output Reading Once
Syntax in STOP-mode: SEND
Syntax in POLL-mode: SEND aa
Where
aa = Address of the transmitter when more than one transmitter is
connected to a serial bus (0 ... 99).
Example:
>send
aw= 0.2 aw T= -47.37 'C …
If the value is too long to fit to the allocated space or if there is an error
in outputting, the value is displayed with stars ‘*’.
For example,
aw=*.* aw T= 31.0 'C
The output mode can be changed with the FORM command.
Set Serial Interface Mode
Syntax: SMODE x
Where
x
= STOP/RUN/POLL
STOP mode: Transmitter in standby for serial
commands
RUN mode: Transmitter outputs data continuously
POLL mode: Transmitter only responds to
addressed commands. Use the OPEN address> command to temporarily enable serial commands (as in STOP mode) when in POLL mode.