Measurement output, Starting the continuous reading output, Stopping the continuous reading output – Vaisala MMT310 User Manual
Page 38: Set continuous output interval for run mode
![background image](/manuals/408548/38/background.png)
User's Guide _______________________________________________________________________
38 ___________________________________________________________________ M210474EN-C
Measurement Output
Starting the Continuous Reading Output
Syntax: R
Starts output of measurements to the peripheral devices (RUN mode).
While the measurement output is active the only command that can be
given is S (stop).
The output mode can be changed with the FORM command.
Example:
>r
aw= 0.2000 aw T= 25.09 'C …
aw= 0.2000 aw T= 25.20 'C …
Stopping the Continuous Reading
Output
Syntax: S
Stops the continuous output.
Set Continuous Output Interval for RUN
Mode
Syntax: INTV xxx yyy
Where
Xxx = Output interval (0 ... 255)
yyy = Unit (s, min or h)
Example:
>intv 1
Output interval: 1 S
>intv 1 min
Output interval: 1 MIN
>intv 1 h
Output interval: 1 H