beautypg.com

Vaisala PTB200 User Manual

Page 27

background image

PTB 200

26 (28)

Digital Barometers

SSD/Operating Manual

24th February 1993

PTB200-O0284-1.1

5

TECHNICAL DATA

OPERATING RANGE

Pressure range

600 ... 1100 hPa

Operating temperature range

-40

°

C ... +60

°

C

Storage temperature range

-60

°

C ... +60

°

C

Humidity range

non-condensing

Note: 1 hPa = 1 mbar (see note below).

ACCURACY

PTB 200 A

PTB 201 A

Linearity *

±0.05 hPa

1)

±0.10 hPa

2)

±0.15 hPa

Hysteresis *

±0.03 hPa

±0.03 hPa

Repeatability *

±0.03 hPa

±0.03 hPa

Calibration uncertainty **

±0.10 hPa

±0.20 hPa

Accuracy at +20

°

C ***

±0.12 hPa

1)

±0.15 hPa

2)

±0.25 hPa

Temperature dependence

±0.1 hPa

±0.1 hPa

Total accuracy (RSS)

±0.20 hPa

±0.3 hPa

Long-term stability

±0.1 hPa / a

+0.2 hPa / a

*

Defined as the ±2 standard deviation limits of end-point non-
linearity, hysteresis error or repeatability error.

**

Defined as ±2 standard deviation limits of inaccuracy of the pri-
mary or working standard at 1000 hPa in comparison to interna-
tional standards (NIST).

***

Defined as the root sum of the squares (RSS) of end-point non-
linearity, hysteresis error, repeatability error and calibration uncer-
tainty at room temperature.

1)

800 to 1100 hPa

2)

600 to 1100 hPa

Note.

The unit hectopascal (hPa) is recommended by WMO to be
used in meteorological barometric pressure measurement (see
WMO Guide to Meteorological Instruments and Methods of Ob-
servation, Fifth edition, 1983).

The official unit of pressure is pascal (Pa) and its multiples
according to the International System of Units (SI) and
according to ISO 1000.