Digital Barometers

SSD/Operating Manual

24th February 1993

PTB200-O0284-1.1

First precalibrate the barometer at up to eight pressure levels. Then
calculate the corrections (reference value minus PTB 200 barometer
indication) at each pressure level and enter the corrections using the
MPCI commands.

>MPCI
1. PTB reading ?499.64
correction ?0.024
2. PTB reading ?599.79
correction ?0.013
...
7. PTB reading ?1100.33
correction ?0.027
8. PTB reading ?

>MPC ON
Multip. corr. : ON
Reading

Corrections

499.6400

0.0240

599.7900

0.0130

.

.

.

.

.

.

1100.3300

0.0270

Use ESC to abort without executing a command.

Return the MEMORY jumper to the EEPROM write disable position.