Vaisala PTB200 User Manual
Page 26

PTB 200
25 (28)
Digital Barometers
SSD/Operating Manual
24th February 1993
PTB200-O0284-1.1
First precalibrate the barometer at up to eight pressure levels. Then
calculate the corrections (reference value minus PTB 200 barometer
indication) at each pressure level and enter the corrections using the
MPCI commands.
>MPCI
1. PTB reading ?499.64
correction ?0.024
2. PTB reading ?599.79
correction ?0.013
...
7. PTB reading ?1100.33
correction ?0.027
8. PTB reading ?
>MPC ON
Multip. corr. : ON
Reading
Corrections
499.6400
0.0240
599.7900
0.0130
.
.
.
.
.
.
1100.3300
0.0270
Use ESC to abort without executing a command.
Return the MEMORY jumper to the EEPROM write disable position.
See also other documents in the category Vaisala Tools:
- DM500 (138 pages)
- DM70 (93 pages)
- DMT132 (74 pages)
- DMT143 (76 pages)
- DMT152 (70 pages)
- DMT242 (4 pages)
- DMT340 (191 pages)
- DMT345 (185 pages)
- DPT145 (63 pages)
- DPT146 (71 pages)
- PTU300 (217 pages)
- PTB330TS (89 pages)
- PTB220 (10 pages)
- PTB220 (113 pages)
- PTB330 (144 pages)
- PTU200 (64 pages)
- PTU200MIK1 (18 pages)
- SPH10 (2 pages)
- SPH20 (2 pages)
- PTB110 (4 pages)
- PTB210 (analog) (27 pages)
- PTB210 (serial) (32 pages)
- GM70 (68 pages)
- GMD20 (4 pages)
- GMK220 (18 pages)
- GML20 (2 pages)
- GML20T (2 pages)
- GMM20W (5 pages)
- GMM220 (6 pages)
- GMP231 (2 pages)
- GMP231 (90 pages)
- GMP343 (94 pages)
- GMR20 (2 pages)
- GMT220 (42 pages)
- GMW90 (101 pages)
- XMW90 (4 pages)
- MM70 (67 pages)
- MM70 (71 pages)
- MMT162 (66 pages)
- MMT310 (81 pages)
- MMT330 (171 pages)
- MMT330 (181 pages)