Vaisala PTB200 User Manual
Page 13

PTB 200
12 (28)
Digital Barometers
SSD/Operating Manual
24th February 1993
PTB200-O0284-1.1
Examples:
>SERI
1200 E 7 1
>SERI 9600 E 7 1
9600 E 7 1
>RESET
>SERI 1200
1200 E 7 1
>RESET
Always give the RESET command after the SERI command to invoke
the new serial bus settings.
The following bus settings do not work with the barometer’s Intel 8051
microprocessor and are modified by the barometer:
N 7 1
⇒
N 7 2
E 8 2
⇒
E 8 1
O 8 2
⇒
O 8 1
Note: N 7 2 works on software version 1.06 or higher only.
The following settings can not be used at all:
E 8 1, O 8 1
ECHO
Setting the serial bus echo on/off
ECHO ON
ECHO OFF
The ECHO command is used to control serial bus echo. In OFF mode
the barometer does not output the ’>’ prompt character.
FORM
Defining the output format
FORM
The FORM command is used to define the desired format and pressure
resolution of the output. Do not use the FORM command to select a
pressure unit (instead see UNIT command).