PTB 200

19 (28)

Digital Barometers

SSD/Operating Manual

24th February 1993

PTB200-O0284-1.1

3.3

Operating commands

R

Starting the measurement output

R

Command R starts the measurement output. The command is used to
start output in the STOP and RUN modes (see SMODE command) and
in interval output mode (see INTV command).

S

Stopping the measurement output

S

Command S ends the RUN mode.

SEND

Output a single reading

SEND

STOP and RUN modes only

SEND aa

POLL mode only

where
aa

the address of the barometer (0...99)

SEND command is used to output one pressure reading.

See also SCOM command to define your own command for single
reading output in the STOP and RUN modes.

INTV

Setting the output interval

INTV xxx yyy

where
xxx

output interval (0...255)

yyy

unit (s, min, h)

INTV command selects the interval output mode and sets the desired
output interval. The R command is used to start the interval outputting.