PTB 200

17 (28)

Digital Barometers

SSD/Operating Manual

24th February 1993

PTB200-O0284-1.1

ADDR

Setting the barometer address (for POLL mode only)

ADDR

ADDR is used to set or inspect an address (1...99) of a barometer for
the POLL mode.

Example of setting the address 7:

>ADDR
Address : 0 ? 7

A new address replaces the previous address. Always set the address
to 0 when no address is needed:

>ADDR
Address : 7 ? 0

If the barometer is not in the POLL mode, it will respond to any SEND
command regardless of if there is an address or not. See SMODE and
SEND commands for further details.

SCOM

Setting a user defined command for single reading
output

(STOP and RUN modes only)

SCOM

is used to define a new command for single reading output in the STOP
and RUN modes. A command defined by SCOM command does not
work in the POLL mode. The SEND command is always available.

Example of a command definition:

>SCOM P
>P
(a single pressure reading appears)

A new SCOM command replaces the previous definition.