Two point calibration, Head – Vaisala HMM213 User Manual

Page 38

background image

USER'S GUIDE____________________________________________________________________

>CT -70 0 101 -0.5
>

Use an accurate and calibrated reference only.

CT Temperature Calibration of the
Additional Sensor Head

Two Point Calibration

CTA

Using this command the transmitters can be calibrated against an
accurate reference, such as a Pt 100 simulator. A two-point calibration is
performed as follows:

>CTA
T : 23.19 Ref1 ?

23.3

Press any key when ready ...
T : 101.42

Ref2 ?

101

If the stabilization of the sensor to the temperature of the reference needs
to be monitored, the measurement output can be repeated with at
Ref1 and Ref2:

>CTA
T : 23.19 Ref1 ?

T : 23.20 Ref1 ?

.
.
.
T : 23.22 Ref1 ?

23.3

Press any key when ready ...
T : 101.42

Ref2 ?

101

38 ____________________________________________________________________ U339EN-1.3