Chapter 5 __________________________________________________________ Serial Commands

Operating via the Serial Bus

SMODE Setting the Serial Interface

SMODE xxxx

xxxx =

STOP, RUN or POLL

In STOP mode:

measurements output only by command, all
commands can be used

In RUN mode:

outputting automatically, only command S can
be used

In POLL mode:

measurements output only with command
SEND. When in POLL mode, the output state is
changed as follows:

OPEN aa

SMODE xxxx

aa =

address of the transmitter

xxxx =

STOP, RUN or POLL

The OPEN command sets the bus temporarily in STOP MODE so that
the SMODE command can be given. For example:

>SMODE

(which mode is in use at the moment)

Output mode : STOP

>SMODE STOP

(setting STOP mode)

Output mode : STOP
>

OPEN & CLOSE

OPEN nn

nn =

address of the transmitter (0 ... 99)

CLOSE

In STOP mode:

command OPEN has no effect, CLOSE sets the
transmitter in POLL mode

In POLL mode:

command OPEN sets the transmitter
temporarily in STOP mode, command CLOSE
returns the instrument to POLL mode