Chapter 5 __________________________________________________________ Serial Commands
Operating via the Serial Bus
SMODE Setting the Serial Interface
SMODE xxxx
xxxx =
STOP, RUN or POLL
In STOP mode:
measurements output only by command, all
commands can be used
In RUN mode:
outputting automatically, only command S can
be used
In POLL mode:
measurements output only with command
SEND. When in POLL mode, the output state is
changed as follows:
OPEN aa
SMODE xxxx
aa =
address of the transmitter
xxxx =
STOP, RUN or POLL
The OPEN command sets the bus temporarily in STOP MODE so that
the SMODE command can be given. For example:
>SMODE
(which mode is in use at the moment)
Output mode : STOP
>SMODE STOP
(setting STOP mode)
Output mode : STOP
>
OPEN & CLOSE
OPEN nn
nn =
address of the transmitter (0 ... 99)
CLOSE
In STOP mode:
command OPEN has no effect, CLOSE sets the
transmitter in POLL mode
In POLL mode:
command OPEN sets the transmitter
temporarily in STOP mode, command CLOSE
returns the instrument to POLL mode