One point calibration, Gain calibration – Vaisala HMM213 User Manual
Page 37

Chapter 6 _______________________________________________________________ Calibration
If the stabilization of the sensor to the temperature of the reference needs
to be monitored, the measurement output can be repeated with
Ref1 and Ref2:
>CTA
T : 23.19 Ref1 ?
T : 23.20 Ref1 ?
.
.
.
T : 23.22 Ref1 ?
23.3
Press any key when ready ...
T : 101.42
Ref2 ?
101
One Point Calibration
CT xx.xx yy.yy
xx.xx =
current temperature
yy.yy =
offset correction
Example of performing the offset calibration in one reference
temperature:
>CT 21.2 2.3
>
Use an accurate and calibrated reference only.
Gain Calibration
CT xx 0 yy.yy zz.zz
xx =
temperature in which there is no changes in offset
yy.yy =
current temperature
zz.zz =
correction (T
ref
-T
measured
)
Example of performing the gain calibration in one reference point
without changing the offset at 0 °C:
>CT 0 0 101 -0.5
>
Example of performing the gain calibration in one reference point
without changing the offset at -70 °C:
VAISALA _______________________________________________________________________ 37