beautypg.com

One point calibration, Gain calibration – Vaisala HMM213 User Manual

Page 37

background image

Chapter 6 _______________________________________________________________ Calibration

If the stabilization of the sensor to the temperature of the reference needs
to be monitored, the measurement output can be repeated with at
Ref1 and Ref2:

>CTA
T : 23.19 Ref1 ?

T : 23.20 Ref1 ?

.
.
.
T : 23.22 Ref1 ?

23.3

Press any key when ready ...
T : 101.42

Ref2 ?

101

One Point Calibration

CT xx.xx yy.yy

xx.xx =

current temperature

yy.yy =

offset correction

Example of performing the offset calibration in one reference
temperature:

>CT 21.2 2.3
>

Use an accurate and calibrated reference only.

Gain Calibration

CT xx 0 yy.yy zz.zz

xx =

temperature in which there is no changes in offset

yy.yy =

current temperature

zz.zz =

correction (T

ref

-T

measured

)

Example of performing the gain calibration in one reference point
without changing the offset at 0 °C:

>CT 0 0 101 -0.5
>

Example of performing the gain calibration in one reference point
without changing the offset at -70 °C:

VAISALA _______________________________________________________________________ 37