Send outputting a reading once, Intv setting the output interval for the run state – Vaisala HMM213 User Manual
Page 22
![background image](/manuals/408437/22/background.png)
USER'S GUIDE____________________________________________________________________
SEND Outputting a Reading Once
SEND
or
SEND aa
aa =
address of the transmitter when more than one transmitter
is connected to a serial bus (0 ... 99; set with command
ADDR)
Outputs the current measurement readings via the serial line. The output
format depends on which parameters the transmitter can output. Output
types are:
"RH 32.25 %RH T 25.74 'C ",
"Td 7.93 'C",
INTV Setting the Output Interval for the
RUN State
INTV xxx yyy
xxx =
output interval (0 ... 255)
0: no pause between outputs
yyy =
unit (s, min or h)
Sets the output interval when the transmitter outputs measurement
readings to a peripheral device.
For example:
>INTV 10 MIN
Output intrv. : 10 min
22 ____________________________________________________________________ U339EN-1.3