SEND Outputting a Reading Once

SEND in STOP state

or

SEND aa in POLL state

aa =

address of the transmitter when more than one transmitter

is connected to a serial bus (0 ... 99; set with command

ADDR)

Outputs the current measurement readings via the serial line. The output
format depends on which parameters the transmitter can output. Output
types are:

"RH 32.25 %RH T 25.74 'C ",
"Td 7.93 'C",

INTV Setting the Output Interval for the
RUN State

INTV xxx yyy

xxx =

output interval (0 ... 255)

0: no pause between outputs

yyy =

unit (s, min or h)

Sets the output interval when the transmitter outputs measurement
readings to a peripheral device.

For example:

>INTV 10 MIN
Output intrv. : 10 min