USER'S GUIDE____________________________________________________________________

When more than one transmitter is connected to the same serial bus, the
POLL mode makes it possible to communicate with the transmitters. For
example, a relative humidity calibration is performed at transmitter 2
( = ASCII 7):

>OPEN 2
'HMM nn line opened for operator commands'

>CRH
...
>CLOSE
'line closed'

Setting the Output Format

MCR Setting the Carriage Return On/Off

MCR x

x =

ON/OFF

>MCR ON
Msg. cr : ON
>

MDEC Selecting a Decimal Separator

MDEC x

x =

a character, TAB, SP, CR, LF

>MDEC .
dec. separator : .
>R
RH 60.50 %RH T 23.90 'C
RH 60.50 %RH T 23.90 'C
...
>MDEC Z
dec. separator : Z
>R
RH 60Z56 %RH T 23Z89 'C
RH 60Z56 %RH T 23Z90 'C
...
>

26 ____________________________________________________________________ U339EN-1.3