TDS Calibration

Press

RANGE to choose the TDS scale.

1. Press

CAL for 2 seconds to enter the calibration mode.

The meter will automatically select the correct range.

The measured value is displayed above the line, the set

calibration value is shown flashing below the line (see Fig. A).

2. Press ↑ once to increase the calibration value.

Hold down ↑ to continually increase the value (see Fig. B).

Press ↓ once to reduce the Calibration value.

Hold down ↓ to continually reduce the value.

3. Press

SETUP to save the value and exit the calibration mode

(see Fig. C).

4. Repeat steps 1-3 to calibrate at other points.

5. Press

EXIT to return to Normal Mode without saving changes.