beautypg.com

Sper Scientific 850034K AquaShock® Water Purity Kit User Manual

Page 12

background image

12

13

Set ATC Temperature Coefficient

1. Press

SETUP to enter the Setup Mode.

2. Press ↑ or ↓ to select the Set Temperature Coefficient screen

(See Fig. A.).

3. Press

SETUP to enter the Set Temperature Coefficient

screen. The current value setting will flash (see Fig. B).

4. Press ↑ once to increase the value. Hold down ↑ to

continually increase the value.

Press ↓ once to decrease the value. Hold down ↓ to

continually decrease the value.

The range may be set from 0.0% to 10.0%.

5. Press

SETUP to save and return to Normal Mode.

6. Press

EXIT to return to Normal Mode without saving change.

To calculate the temperature compensation coefficient:

1. Set temperature coefficient to 0%.

2. Put conductivity probe into the standard calibration solution.

3. After 5 minutes record the first temperature reading and the

corresponding conductivity value.

4. Heat or cool the solution. After a change of 5° to 10°C record

the second temperature reading along with the corresponding

conductivity value.

5. Calculate temperature coefficient by formula below:

Fig. A

Fig. B

TC = temperature coefficient

CT1 = conductivity value when temperature is 1

CT2 = conductivity value when temperature is 2

T1 = temperature value 1

T2 = temperature value 2

25 = 25°C