Set TDS Factor

1. Press

SETUP to enter the Set up mode.

2. Press ↑ or ↓ to select the Set TDS Factor screen (see Fig. A).

3. Press

SETUP to enter the mode, the current value setting

will flash (see Fig. B).

4. Press ↑ once to increase the value. Hold down ↑

to continually increase the value.

Press ↓ once to decrease the value. Hold down ↓

to continually decrease the value.

The range may be set from 0.40 to 1.00.

5. Press

SETUP once to confirm the TDS factor value

and return to Normal Mode.

6. Press

EXIT once to return to Normal Mode without

saving changes.

Fig. A

Fig. B

Cond.

value at

25°C

TDS KCl

TDS NaCl

TDS 442

TDS Your

Material

ppm

Factor ppm

Factor

ppm

Factor ppm

Factor

84 μS

40.38

0.5048 38.04

0.4755 50.50

0.6563

447 μS

225.6

0.5047 215.5

0.4822 300.0

0.6712

1413 μS

744.7

0.5270 702.1

0.4969 1000

0.7078

1500 μS

757.1

0.5047 737.1

0.4914 1050

0.7000

8974 μS

5101

0.5685 4487

0.5000 7608

0.8478

12880 μS 7447

0.5782 7230

0.5613 11,367 0.8825

15000 μS 8759

0.5839 8532

0.5688 13,455 0.8970

80 mS

52,168 0.6521 48,384 0.6048 79,688 0.9961