CALBRATION
Before calibration, place the probe in the standard calibration

solution. If measured values exceed the standard solution

concentration by ±20%, re-clean your probe or replace it.

Conductivity Calibration

Press

RANGE to choose the conductivity scale.

1. Press

CAL for 2 seconds to enter the calibration mode.

The meter will automatically select the correct range.

The measured value is displayed above the line, the set

calibration value is shown flashing below the line (see Fig. A).

2. Press ↑ once to increase the Calibration value.

Hold down ↑ to continually increase the value.

Press ↓ once to reduce the Calibration value.

Hold down ↓ to continually reduce the value (see Fig. B).

3. Press

SETUP to save the value and exit the calibration

mode (see Fig. C).

4. Repeat steps 1-3 to calibrate at other points.

5. Press

EXIT to return to Normal Mode without saving changes.