Problem

Probable Cause

Solution

Condensation collects in the
oxygen tubing when you use
a humidifier bottle

Unit not properly
ventilated.
Elevated operating
temperatures.

Make sure unit is
positioned away from
curtains or drapes, hot
air registers or heaters,
Be certain to place the
unit so all sides are at
least 1 foot (30 cm)
away from a wall or
other obstruction. Do
not place the unit in a
confined area.
Allow oxygen tubing to
dry out, or replace with
new tubing. Refill
humidifier bottle with
COLD water.
DO NOT OVERFILL.

Intermittent alarm sounds at
one-second intervals.

Equipment malfunction.

Set I/0 power switch to
0 position, use your
reserve oxygen supply
(if provided), and
consult your Equipment
Provider immediately.

All other problems.

Set I/0 power switch to
the 0 position, use your
reserve oxygen supply
(if provided), and
consult your Equipment
Provider immediately.