beautypg.com

AirSep MN136-1 H User Manual

Page 7

background image

[Read the Important Safety Rules section before operating this equipment.]

4 MN136-1 rev H


The concentrator should be located as to avoid pollutants or fumes.


If oxygen does not seem to flow, first verify that the flowmeter ball is
registering a flow. Then, place the tip of the cannula into a glass of water; if
bubbles come out of the cannula, oxygen is flowing. If bubbles do not
appear, refer to Section 5.0. Troubleshooting.

To prevent a voided AirSep warranty, follow all manufacturers’
instructions.


Replace the disposable cannula periodically following normal usage
according to your Equipment Providers recommendations.


AirSep does not recommend the sterilization of this equipment.


Do not attempt any maintenance other than the possible solutions listed
below.


If the unit has not been used for an extended time period, it needs to operate
for several minutes before power failure alarm can become activated.

Portable and mobile RF communications equipment can affect medical
electrical equipment.

The concentrator releases warm air out the bottom of the unit which can
permanently discolor temperature sensitive flooring surfaces such as vinyl.
The concentrator should not be used over flooring that is sensitive to heat
staining. AirSep is not responsible for flooring that becomes discolored.

“No Smoking - Oxygen in Use” signs must be prominently displayed in the
home or where the oxygen concentrator is in use. Patients and their
caregivers must be informed about the dangers of smoking in the presence
of, or while using, medical oxygen.