40 Calibration
500
INPUT "ENTER VALUE OF CURRENT SHUNT BEING USED",Shunt_val
510
OUTPUT @Ps;"CAL:CURRENT:LEVEL MIN"
520
INPUT "ENTER VOLTAGE MEASUREMENT FROM EXTERNAL VOLTMETER",Volt_read
530
Current=Volt_read/Shunt_val
540
OUTPUT @Ps;"CAL:CURRENT ";Current
550
OUTPUT @Ps;"CAL:CURRENT:LEVEL MAX"
560
INPUT "ENTER VOLTAGE MEASUREMENT FROM EXTERNAL VOLTMETER",Volt_read
570
Current=Volt_read/Shunt_val
580
OUTPUT @Ps;"CAL:CURRENT ";Current
590
GOSUB
Save_cal
600
IF Err_found THEN
610
INPUT "ERRORS have occurred, REPEAT CURRENT
CALIBRATION (Y OR N)?",Resp$
620
IF TRIM$(UPC$(Resp$[ l,l ]))="Y" THEN GOTO
Current_cal
630
END
IF
640
IF Err_found THEN
650
PRINT "CURRENT CALIBRATION NOT SAVED"
660
ELSE
670
PRINT "CURRENT CALIBRATION COMPLETE"
680
END
IF
690
STOP
700 Save_cal:
! SAVE CALIBRATION CONSTANTS
710
REPEAT
720
OUTPUT @Ps;"SYSTEM:ERROR?"
730
ENTER @Ps;Err_num,Err_msg$
740
IF Err_num<>0 THEN
750
PRINT "ERROR: ";Err_msg$
760
Err_found=l
770
END IF
780
UNTIL Err_num=0
790
IF NOT Err_found THEN
800
INPUT "SAVE CALIBRATION CONSTANTS (Y OR N)?",Resp$
810
IF TRIM$(UPC$(Resp$[1,1]))="Y" THEN
820
OUTPUT @Ps;"CAL:SAVE"
830
END IF
840
END IF
850
OUTPUT @Ps;"CAL:STATE 0"
860
RETURN
870
END
Figure A-2. Agilent BASIC Calibration Program (continued)