When the power module is operated beyond either rated output, the performance specifications are not guaranteed, although
typical performance may be good. Operation in quadrant 2 is limited to about 10% of the maximum rated positive output
current. This allows the power module to sink current for more rapid downprogramming in the CV mode.

Specifications and Supplemental Characteristics

Table 1-1 lists the specifications of the power modules. Specifications are warranted over the specified temperature range.
Table 1-2 lists the supplemental characteristics, which are not warranted but are descriptions of performance determined
either by design or type testing.

Table 1-1. Performance Specifications

Agilent Model

Parameter

66101A

66102A

66103A

66104A

66105A

66106A

Specifications are warranted over the temperature range 0 to 55

°

C with a resistive load and the output connected for local

sensing. Derate the output current 1% per

°

C from 40

°

C to 55

°

C.

Output Ratings
Voltage:

0 - 8 V

0 - 20 V

0 - 35 V

0 - 60 V

0 - 120 V

0 - 200 V

Current:

0 - 16 A

0 - 7.5 A

0 - 4.5 A

0 - 2.5 A

0 - 1.25 A

0 - 0.75 A

Power:

128 W

150 W

157.5 W

150 W

150 W

150 W

Programming Accuracy @
calibration temperature ± 5

°

C)*

Voltage: 0.03% +

3 mV

8 mV

13 mV

27 mV

54 mV

90 mV

Current: 0.03% +

6 mA

3 mA

2 mA

1.2 mA

0.6 mA

0.4 mA

*Factory calibration temperature =
25

°

C

Ripple & Noise (from 20 Hz to 20
MHz with outputs ungrounded, or with
either output terminal grounded)

Constant Voltage:

rms

2 mV

3 mV

5 mV

9 mV

18 mV

30 mV

Constant Voltage: p-p

5 mV

7 mV

10 mV

15 mV

25 mV

50 mV

Constant Current: rms

8 mA

4 mA

2 mA

1 mA

1 mA

1 mA

Readback Accuracy
With respect to actual output @
calibration temperature following self
calibration.*
Voltage:

0.02% +

2 mV

5 mV

8 mV

16 mV

32 mV

54 mV

Current:

0.02% +

6 mA

3 mA

2 mA

1 mA

0.6 mA

0.3 mA

*Factory calibration temperature =
25

°

C

Load Regulation (change in output
voltage or current for any load change
within ratings)

Voltage:

1 mV

1 mV

1 mV

2 mV

4 mV

7 mV

Current:

0.5 mA

0.2 mA

0.2 mA

0.1 mA

50

µ

A

30

µ

A