Agilent Technologies 66001A User Manual
Page 12
12 General Information
Table 1-1. Performance Specifications (continued)
Agilent Model
Parameter
66101A
66102A
66103A
66104A
66105A
66106A
Line Regulation (change in output
voltage or current for any line change
within ratings)
Voltage:
0.5 mV
0.5 mV
1 mV
2 mV
3 mV
5 mV
Current:
0.75 mA
0.5. mA
0.3 mA
0.1 mA
50
µ
A
30
µ
A
Transient Response Time (for output
voltage to recover to within 100 mV of
its previous level following a step
change in load current that is up to 10%
of the power module rated output
current.
< 1 ms
AC Input Ratings (selectable via
module switch- see Chapter 2)
Nominal line voltage (Vac)
115 V switch setting
87- 132 Vac
230 V switch setting
174- 250 Vac
Frequency range:
47 - 63 Hz
Output Terminal Isolation (maximum,
from output to the mainframe chassis
ground):
±240 Vdc
Table 1-2. Supplemental Characteristics
Agilent Model
Parameter
66101A
66102A
66103A
66104A
66105A
66106A
Output Programming Range
(Maximum programmable values):
Voltage:
8.190 V
20.475 V
35.831 V
61.425 V
122.85 V
204.75 V
Current:
16.380 A
7.678 A
4.607 A
2.559 A
1.280 A
0.768 A
OV Protection:
9.6 V
24.0 V
42.0 V
72.0 V
144.0 V
240.0 V
Average Programming Resolution
Voltage:
2.4 mV
5.9 mV
10.4 mV
18.0 mV
36.0 mV
60.0 mV
Current:
4.6 mA
2.3 mA
1.4 mA
0.75 mA
0.39 mA
0.23 mA
OV Protection:
50 mV
120 mV
200 mV
375 mV
750 mV
1.25 V
OVP Accuracy (@ calibration
temperature ±5
°
C):*
250 mV
500 mV
800 mV
1 V
1.5 V
2.5 V
* Factory calibration temp = 25
°
C
Readback Resolution
Voltage:
0.305 mV
0.763 mV
1.335 mV
2.289 mV
4.577 mV
7.629 mV
Current:
0.587 mA
0.293 mA
0.169 mA
0.099 mA
0.049 mA
0.030 mA
Drift (Temperature Stability)
(following a 30-minute warmup, change
in output over 8 hours under constant
line, load, and ambient temperature)
Voltage:
0.02% +
0.4 mV
1 mV
2 mV
3 mV
6 mV
10 mV
Current:
0.02% +
8 mA
5 mA
2.5 mA
1.3 mA
0.63 mA
0.38 mA