Voltage Configuration Commands

(see page 223 in the User’s Guide)

CONFigure
:VOLTage:DC [{<

range

>|

AUTO

|MIN|MAX|DEF}

[,<

resolution

>|MIN|MAX|DEF}],] (

@

<

scan_list

>)

CONFigure? [(

@

<

ch_list

>)]

[SENSe:]
VOLTage:DC:RANGe {<

range

>|MIN|MAX}[,(

@

<

ch_list

>)]

VOLTage:DC:RANGe? [{(

@

<

ch_list

>)|MIN|MAX}]

VOLTage:DC:RANGe:AUTO {OFF|

ON

}[,(

@

<

ch_list

>)]

VOLTage:DC:RANGe:AUTO? [(

@

<

ch_list

>)]

[SENSe:]
VOLTage:DC:RESolution {<

resolution

>|MIN|MAX}[,(

@

<

ch_list

>)]

VOLTage:DC:RESolution? [{(

@

<

ch_list

>)|MIN|MAX}]

[SENSe:]
VOLTage:DC:APERture {<

time

>|MIN|MAX}[,(

@

<

ch_list

>)]

VOLTage:DC:APERture? [{(

@

<

ch_list

>)|MIN|MAX}]

[SENSe:]
VOLTage:DC:NPLC {0.02|0.2|

1

|2|10|20|100|200|MIN|MAX}[,(

@

<

ch_list

>)]

VOLTage:DC:NPLC? [{(

@

<

ch_list

>)|MIN|MAX}]

INPut
:IMPedance:AUTO {

OFF

|ON}[,(

@

<

ch_list

>)]

:IMPedance:AUTO? [(

@

<

ch_list

>)]

[SENSe:]
ZERO:AUTO {OFF|ONCE|

ON

}[,(

@

<

ch_list

>)]

ZERO:AUTO? [(

@

<

ch_list

>)]

CONFigure
:VOLTage:AC [{<

range

>|

AUTO

|MIN|MAX|DEF}

[,<

resolution

>|MIN|MAX|DEF}],] (

@

<

scan_list

>)

CONFigure? [(

@

<

ch_list

>)]

[SENSe:]
VOLTage:AC:RANGe {<

range

>|MIN|MAX}[,(

@

<

ch_list

>)]

VOLTage:AC:RANGe? [{(

@

<

ch_list

>)|MIN|MAX}]

VOLTage:AC:RANGe:AUTO {OFF|

ON

}[,(

@

<

ch_list

>)]

VOLTage:AC:RANGe:AUTO? [(

@

<

ch_list

>)]

[SENSe:]
VOLTage:AC:BANDwidth {3|

20

|200|MIN|MAX}[,(

@

<

ch_list

>)]

VOLTage:AC:BANDwidth? [{(

@

<

ch_list

>)|MIN|MAX}]