Temperature Configuration Commands

(see page 219 in the User’s Guide)

CONFigure
:TEMPerature {

TCouple

|RTD|FRTD|THERmistor|DEF}

,{<

type

>|DEF}[,1[,{<

resolution

>|MIN|MAX|DEF}]] ,(

@

<

scan_list

>)

CONFigure? [(

@

<

ch_list

>)]

UNIT
:TEMPerature {

C

|F|K}[,(

@

<

ch_list

>)]

:TEMPerature? [(

@

<

ch_list

>)]

[SENSe:]TEMPerature:TRANsducer
:TYPE {

TCouple

|RTD|FRTD|THERmistor|DEF}[,(

@

<

ch_list

>)]

:TYPE? [(

@

<

ch_list

>)]

[SENSe:]TEMPerature:TRANsducer
:TCouple:TYPE {B|E|J|

K

|N|R|S|T}[,(

@

<

ch_list

>)]

:TCouple:TYPE? [(

@

<

ch_list

>)]

:TCouple:CHECk {

OFF

|ON}[,(

@

<

ch_list

>)]

:TCouple:CHECk? [(

@

<

ch_list

>)]

[SENSe:]TEMPerature:TRANsducer
:TCouple:RJUNction:TYPE {

INTernal

|EXTernal|FIXed}[,(

@

<

ch_list

>)]

:TCouple:RJUNction:TYPE? [(

@

<

ch_list

>)]

:TCouple:RJUNction {<

temperature

>|MIN|MAX}[,(

@

<

ch_list

>)]

:TCouple:RJUNction? [(

@

<

ch_list

>)]

[SENSe:]TEMPerature:RJUNction? [(

@

<

ch_list

>)]

[SENSe:]TEMPerature:TRANsducer
:RTD:TYPE {

85

|91}[,(

@

<

ch_list

>)]

:RTD:TYPE? [(

@

<

ch_list

>)]

:RTD:RESistance[:REFerence] <

reference

>[,(

@

<

ch_list

>)]

:RTD:RESistance[:REFerence]? [(

@

<

ch_list

>)]

[SENSe:]TEMPerature:TRANsducer
:FRTD:TYPE {

85

|91}[,(

@

<

ch_list

>)]

:FRTD:TYPE? [(

@

<

ch_list

>)]

:FRTD:RESistance[:REFerence] <

reference

>[,(

@

<

ch_list

>)]

:FRTD:RESistance[:REFerence]? [(

@

<

ch_list

>)]

[SENSe:]TEMPerature:TRANsducer
:THERmistor:TYPE {2252|

5000

|10000}[,(

@

<

ch_list

>)]

:THERmistor:TYPE? [(

@

<

ch_list

>)]

[SENSe:]
TEMPerature:NPLC {0.02|0.2|

1

|2|

10

|20|100|200|MIN|MAX}[,(

@

<

ch_list

>)]

TEMPerature:NPLC? [{(

@

<

ch_list

>)|MIN|MAX}]

This command redefines the scan list when executed.

Default parameters are shown in bold.

5