Eagle s, Pecifications – UEi Test Instruments C127OILKIT User Manual
Page 19

EAGLE S
PECIFICATIONS
18
19
Features
C75
C125
C127
Worklight
4
4
Low battery indicator
4
4
Individual print-out reports
4
4
Rotary selector
4
4
4
Memory positions (various tests)
20
179
179
Infrared printer port
4
4
4
Backlit display
2 Lines
4 Lines
4 Lines
Boot with integral magnet
4
4
4
User programmable headers
4
4
4
Time & date stamp
4
4
4
Temp Measurement
Flue Temp Range (T1 C125/C127)
32 ~ 1112˚F / 0 ~ 600˚C
20 ~ 2400˚F / -29 ~ 1315˚C
Inlet Temp Range:Probe (T2 C125/C127)
-
Inlet Temp Range: Ambient
32 ~ 112˚F / 0 ~ 50˚C
32 ~ 112˚F / 0 ~ 50˚C
Net Temp (∆T)**
32 ~ 1112˚F / 0 ~ 600˚C
20 ~ 2400˚F / -29 ~ 1315˚C
Resolution
1˚F/˚C
0.1˚F/˚C
Flue (T1, Inlet T2, & Net ∆T) Accuracy
±[0.3% rdg +5˚F(3˚C)]
±[0.3% rdg +3.6˚F(2˚C)]
Inlet Temp Ambient Accuracy
±[0.3% rdg +1˚F(1˚C)]
±[0.3% rdg +1.8˚F(1˚C)]
Gas Measurement
Oxygen
0 ~ 21%
0 ~ 21%
0 ~ 21%
O2 Resolution/Accuracy
0.1% / ±0.2%*
0.1% / ±0.2%*
0.1% / ±0.2%*
Carbon Monoxide
0 ~ 1000 ppm
0 ~ 4000 ppm
0 ~ 4000 ppm
CO Resolution/Accuracy
1 ppm/±10 ppm < 100 ppm*
1 ppm/±10 ppm < 100 ppm*
1 ppm/±10 ppm < 100 ppm*
±5% reading
±5% reading
±5% reading
Carbon Dioxide**
0 ~ 30%
0 ~ 30%
0 ~ 30%
CO2 Resolution/Accuracy
1.0%/±0.3% reading
1.0%/±0.3% reading
1.0%/±0.3% reading
Efficiency**
0 ~ 99.9%
0 ~ 99.9%
0 ~ 99.9%
Efficiency Resolution/Accuracy
0.1%/±1.0% reading
0.1%/±1.0% reading
0.1%/±1.0% reading
Excess Air**
0 ~ 250%
0 ~ 250%
0 ~ 250%
Excess Air Resolution/Accuracy
0.1%/±0.2% reading
0.1%/±0.2% reading
0.1%/±0.2% reading
CO/CO2**
-
0 ~ 0.999%
0 ~ 0.999%
CO/CO2 Resolution/Accuracy
-
0.0001%/±5% reading
0.0001%/±5% reading
Nitric Oxide
-
-
1 ppm/±2 ppm < 30 ppm*
Nitric Oxide Resolution/Accuracy
-
-
±5 ppm > 30 ppm
Pressure (Differential)
Range ±0.2mBar 0.08”wg
0.001mBar/±0.005mBar ±0.002”wg
0.001mBar/±0.005mBar ±0.002”wg
Range ±1mBar 0.4”wg
0.001mBar/±0.03mBar ±0.01”wg
0.001mBar/±0.03mBar ±0.01”wg
Range ±80mBar 32”wg
0.01mBar/±3% reading
0.01mBar/±3% reading
Pre-Programmed Fuels
Natural Gas, Propane, Butane, LPG, Light Oil
Eagle Dimensions
Weight
2.2lbs. with boot
Handset
7.9 x 1.8 x 3.5”
Probe
11.8 x 0.25 x 9.4” stainless steel shaft, type K thermocouple (9’ hose)
Operating temp
0 to 40˚C/10% to 90% Relative Humidity non-condensing
Battery
x4 “AA” cells / 12 hour life typical with Alkaline “AA” cells
Warranty
3 year limited warranty (2 year sensor warranty)
*Measures **Calculates