TOHO ELECTRONICS TTM-10L User Manual
Page 7
ORDERING INFORMATION
8
8
8
8.
.
.
.
TTM−10L−□−□−A□
記号
記号
記号
Input
Output
Option
0
N
Communication
M1
Thermocouple / R.T.D.
None
4 ∼
20mA DC
1
R
RS−485
Relay contact
P
Communication
M2
SSR drive volt.
RS−232C
SPECIFICATIONS
9
9
9
9.
.
.
.
9.1 GENERAL SPECIFICATION(COMMON)
Power Supply Voltage
12VDC±10%
or
24VDC±10%
Power Consumption
12VDC:below 1.4W
24VDC:below 2.7W
Operation
Temperature
0∼50℃
/
35∼85%RH (Avoid making dew)
environment
& Humidity
range
Supply
12VDC±10%
or
24VDC±10%
Voltage
Transportation
Temperature
−20∼60℃
/
35∼85%RH (Avoid making dew)
storage
& Humidity
environment
range
Drop from the
Vibration
0.5G
Impact
0∼50G
Package
Condition
Condition
Drop test
60cm
height
Mechanical
Weight
TTM-10L less then 80g
Specifications
9.2 GENERAL SPECIFICATION(DETAILS)
Switchable.
K、J、E、T、N、R、S、W5Re/W26Re
Thermocouple
Input resistance: more than 1MΩ
Bias Current:Approx 0.2μV/Ω
Burn out:"OVER"(
)
Switchable.
PV INPUT
Kind of
Pt100/JPt100
INPUT
R.T.D.
External resistance:
5Ω(Per wire)
less than
Burn out:Shown"OVER"(
)
same display shown at snapping of either A,B or b.
CURRENT
Input resistance:250Ω
(4-20mA DC)
Burn out:Shown"UNDER"(
)
Sampling
cycle
0.5sec(same as Output alteration cycle)
PV Display
4figures
7segmentLCD
Letter height 7mm
Character Display
1figure
7segmentLCD
Letter height 5mm
PV/Character
Control Output
OUT Lamp
Light ON at Control Output
Display
Alarm Output
AL Lamp
Light ON at Alarm Output
Communication
COM Lamp
Light ON at operating Communication
Thermocouple
±0.5%+1digit of indicated value,or±4℃(8),
(
i on of
Stipulated by the standard environment and the convers
Indication
indicated value)
Accuracy
R.T.D.
±0.5%+1digit of indicated value, or±1.2℃(2.4),
(
i on of
Stipulated by the standard environment and the convers
indicated value)
±0.5%+1digit of Setting Value Limiter Span.
CURRENT (4-20mA DC)
(
i on of
Stipulated by the standard environment and the convers
indicated value)
All Output functions being OFF for approximately 4 seconds after Power ON.
Output at
Control Output being OFF at abnormal Process Value.
Special cases
Contact specification :1a Contact point.
Contact capacity :250VAC 3A(Load resistance)
Control &
Relay Contact
Output
Kind of
Output
Endurance :more than 100,000 times.
Section
Output and
SSR Drive
Output Voltage :OFF time;0V
ON time ;12V+10% -20%
Rating
Voltage Output
Loaded resistance :0ver 600 Ω
Alarm
Contact specification :1a Contact point.
Contact capacity :250VAC 0.5A(Load resistance)or 125VAC 1A(Load resistance)
Functions
Alarm Output
Endurance :more than 100,000 times.