a) Procedure for adjusting the probe using the optional keypad (HF53 / HF55)

Temperature adjustment

The keypad of the HF53 or HF55 allows only a 1-point adjustment of temperature. The effect of a 1-point
temperature adjustment is to add the same offset to all measured values.

When temperature is stable, press the MENU key to show the internal menu on the display
Use the (-) key to select Temperature Adjust and press the ENTER key
Make sure that the text line beginning with RefValue is highlighted and press the ENTER key
Use the (+) or (-) key to change the reference value to match the temperature reference
Use the (-) key to highlight the Adjust text line and press the ENTER key

The HF5 confirms the adjustment with the message “Adjust OK”

Press the MENU key twice to exit the menu and return the HF5 to normal operation


Notes:

o

Any temperature adjustment should be done prior to adjusting humidity.

o

The calibration point is automatically deleted from the probe memory after an adjustment

o

Because the HF5 has no real time clock, the date of the adjustment is not written to the probe. If
retaining the adjustment date is important, use the HW4 software to adjust the probe.


Humidity adjustment

The keypad of the HF53 or HF55 allows a multi-point adjustment of humidity. The effect of a humidity
adjustment depends on the number of calibration points present in the probe memory prior to the
adjustment:

- one calibration point: offset added to all measured values
- two calibration points: offset and slope
- three or more calibration points: offset, slope, linearity


For maximum accuracy use at least 3 to 4 calibration points distributed equally across the measurement
range of interest. The calibration points (maximum 100) can be acquired in any order but we recommend
going from low humidity values to high humidity values.

When humidity is stable, press the MENU key to show the internal menu on the display
Use the (-) key to select Humidity Adjust and press the ENTER key
Make sure that the text line beginning with RefValue is highlighted and press the ENTER key
Use the (+) or (-) key to change the reference value to match the reference humidity
Use the (-) key to highlight the Acquire text line and press the ENTER key

Note that the “Acquired” counter is incremented by 1 (number of calibration points in the probe
memory)

When all calibration points have been acquired, use the (-) key to highlight the Adjust text line and

press the ENTER key. Do not adjust the probe before having acquired all calibration points.

The HF5 confirms the adjustment with the message “Adjust OK”

Press the MENU key twice to exit the menu and return the HF5 to normal operation


Notes:

o

All calibration points are automatically deleted from the probe memory after an adjustment

o

Instructions for using the ROTRONIC calibration devices and humidity standards are provided in
document E-M-CalBasics

o

Because the HF5 has no real time clock, the date of the adjustment is not written to the probe. If
retaining the adjustment date is important, use the HW4 software to adjust the probe.