beautypg.com

Dwyer instruments, inc, Caution – Dwyer VFB User Manual

Page 2

background image

the boot. Pipe up inlet and discharge using pipe thread sealant tape or

pipe thread sealant to insure against leakage.

Surface Mounting on Piping Only: An alternate method of surface

mounting is by omitting the mounting screws and supporting the Visi-

Float

®

Flowmeter on the connecting piping only. For this method extra

long or straight pipe threads should be used so that nuts may be run onto

the pipe and later tightened against the back of the panel to retain the

unit in proper position. Use the appropriate hole layout information from

Figure 1, but omit the small holes.

Mounting on Piping Only Without Panel: For the temporary or

laboratory type installation, the panel may be omitted altogether and the

flowmeter installed directly in rigid piping. Its light weight permits this

without difficulty.

OPERATION

To start system, open the valve slowly to avoid possible damage. Control

valves on BV and SSV models are turned clockwise to reduce flow,

counter clockwise to increase flow (valve is designed for flow adjustment

only, not intended to be used as an open/shut-off valve). A nylon insert is

provided in the threaded section of the valve stem to give a firm touch to

the valve and to prevent change of setting due to vibration.

The performance of low range units used in air or gas applications may

be affected by static electricity. Excessive static charge may cause the

ball float to behave erratically or provide a false reading. To ensure the

proper function of the unit, the application should be designed to

minimize or dispel static electricity.

The standard technique for reading a Variable Area Flowmeter is to

locate the highest point of greatest diameter on the float, and then align

that with the theoretical center of the scale graduation. In the event that

the float is not aligned with a grad, an extrapolation of the float location

must be made by the operator as to its location between the two closest

grads. The following are some sample floats shown with reference to the

proper location to read the float.

Variable Area Flowmeters used for gases are typically labeled with the

prefix “S” or “N”, which represents “Standard” for English units or

“Normal” for metric units. Use of this prefix designates that the flowmeter

is calibrated to operate at a specific set of conditions, and deviation from

those standard conditions will require correction for the calibration to be

valid. In practice, the reading taken from the flowmeter scale must be

corrected back to standard conditions to be used with the scale units.

The correct location to measure the actual pressure and temperature is

at the exit of the flowmeter, except under vacuum applications where

they should be measured at the flowmeter inlet. The equation to correct

for nonstandard operating conditions is as follows:

Q

2

= Q

1

x P

1

x T

2

P

2

x T

1

Where:

Q

1

= Actual or Observed Flowmeter Reading

Q

2

= Standard Flow Corrected for Pressure and

Temperature

P

1

= Actual Pressure (14.7 psia + Gage Pressure)

P

2

= Standard Pressure (14.7 psia, which is 0 psig)

T

1

= Actual Temperature (460 R + Temp °F)

T

2

= Standard Temperature (530 R, which is 70°F)

Example: A flowmeter with a scale of 10-100 SCFH Air. The float is sitting

at the 60 grad on the flowmeter scale. Actual Pressure is measured at

the exit of the meter as 5 psig. Actual Temperature is measured at the

exit of the meter as 85°F.

Q

2

= 60.0 x (14.7 + 5) x 530

14.7 x (460 + 85)

Q

2

= 68.5 SCFH Air

MAINTENANCE

The only maintenance normally required is occasional cleaning to assure

reliable operation and good float visibility.

Disassembly: The flowmeter can be disassembled for cleaning by

simply disconnecting the piping, dismounting the unit from the panel and

removing the top-plug-ball stop. Take out the ball or float by inverting the

body and allowing the float to fall into your hand. (Note: It is best to cover

the discharge port to avoid losing the float through that opening.) When

removing the float guide assembly on VFB models 85 and 86, be careful

not to lose the short pieces of plastic tubing on both ends of the guide

which serve as float stops.

Cleaning: The flow tube and flowmeter body can best be cleaned with a

little pure soap and water. Use of a bottle brush or other soft brush will

aid the cleaning. Avoid benzene, acetone, carbon tetrachloride, alkaline

detergents, caustic soda, liquid soaps (which may contain chlorinated

solvents), etc. and avoid prolonged immersion.

Reassembly: Reinstall the float, remount, connect and place the unit

back in service. A little stop cock grease or petroleum jelly on the “O”rings

will help maintain a good seal as well as facilitate assembly. No other

special care is required.

For VFB models 85 and 86, first install the lower fitting, next the float

guide and float and finally the upper fitting and plug. Be certain that both

ends of the float guide are properly engaged and the float is correctly

oriented.

ADDITIONAL INFORMATION

For additional flowmeter application information, conversion curves,

factors and other data covering the entire line of Dwyer Instruments, Inc.

full-line catalog.

Do not completely unscrew valve stem unless flowmeter is
unpressurized and drained of any liquid. Removal while in service
will allow gas or liquid to flow out front of valve body and could
result in serious personal injury. For applications involving high
pressure and/or toxic gases or fluids, special non-removable
valves are available on special order. Contact factory for details.

CAUTION

Model VFA — 2" Scale

Range

SCFH Air

.1-1
.2-2
.6-5

1-10
2-20
4-30
5-50

10-100
20-200

CC Water

per min.

6-50

10-100
20-200

Range

No.

1
2
3
4
5
6
7
8
9

32
33
34

Range

LPM Air

.06-0.5

.15-1

.6-5

1-10
3-25
6-50

10-100

Gal. Water

per hour

.6-5

1.2-10

3-20
8-40

Range

No.

21
22
23
24
25
26
27

41
42
43
44

Model VFB — 4" Scale

Range

SCFH Air

.3-3

1-10
2-20
4-40

10-100
15-150
20-200

SCFM Air

.3-3

CC/Min. Air

100-1000

Range

No.

50
91
51
52
53
54
55

90

60

LPM Air

.2-4

1-10
1-20
3-30
4-40

CC/Min. Water

2-30

GPH Water

.5-12

1-20
6-60

GPM Water

.2-2
.5-5

Range

No.

65
66
67
68
69

82

80
83
81

85
86

POPULAR RANGES

©Copyright 2013 Dwyer Instruments, Inc.

Printed in U.S.A. 6/13

FR# 52-440241-00 Rev.6

DWYER INSTRUMENTS, INC.

Phone: 219/879-8000

www.dwyer-inst.com

P.O. BOX 373 • MICHIGAN CITY, IN 46360, U.S.A.

Fax: 219/872-9057

e-mail: [email protected]

This manual is related to the following products: