4 functions, Operating modes, 5 menu – Dwyer HM35 User Manual

Page 7: Selection and set-up, 4 functions and operating modes

background image

3.4

Functions and operating modes

Key

clear

hold

1)

Start

2)

Functions

PRESSURE

DIFFERENCE

Zero: sets
measured value to
zero

3)

MIN/MAX

Sets Max/Min to
current measured
value

To Menu selection

--

AVERAGE

4)

CHANGE RATE

5)

Freezes all

current

measurement

values

Starts measuring

DATA LOGGING

Sets measured
value to zero

3)

--

Stop/Menu

Starts data
logging


Notes

1) Once a measurement procedure has been started, the menu selection is blocked.
2) During a measurement procedure (after Start has been pressed), you can switch

between functions. This permits, for example, the observation of the Min/Max
function during data logging.

3) The Clear key has no function in the absolute pressure instruments.
4) The AVERAGE function creates an arithmetic average value of all measured values

during the time period selected in the menu. After expiry of the time period, the
average value will be displayed.

5) Measurement

of

the leak rate (diff/gauge sensor) or tendency (abs. sensor). The

pressure change (CHANGE RATE) from the start time to the current time will be
displayed. The first display occurs 10 s after the start.



3.5

Menu selection and set-up


Navigation within the Menu selection

Brief (< 1 s)

1 level back

clear

Long (> 1 s)

Back to the function level/operating modes

▲▼ Selection

of

Set-up/Functions


The functions shown inverted on the display will be carried out if the Function key
(

) is pressed.

The currently selected setting for values is marked with '9'. In the following table, the
default values are correspondingly marked (factory settings).

6