8 battery, Replacement, 9 calibration – Dwyer HM35 User Manual

Page 20

background image

3.8 Battery

replacement


• Open the battery compartment and insert 3 x 1,5 V Mignon cell AA, IEC LR6.

Always replace all three batteries at the same time!

Ensure correct polarity!

Dispose of used batteries in accordance with environmental
regulations

!




3.9 Calibration

Re-calibration may only be carried out by specialist staff and with the corresponding
pressure standards.
We recommend that you have the HM35 re-calibrated at least once a year, and, in case
of highest demands for precision, several times a year.

Manual re-calibration

• In

the

Menu,

select the Calibration

Manual Calibration function.

Zero point

(Offset)

1

. Open the pressure connection or, with the absolute pressure unit, set the given

pressure value to the normal pressure.

2.

Press

the

Function

key (

)

Î

the zero point will be re-calibrated.



Full-scale value

1

.

Set the given pressure value to the normal pressure.

2

. Press

the

Function

key (

)

Î

the full-scale value will be calibrated and the HM35 returns to normal operation.

19