8 battery, Replacement, 9 calibration – Dwyer HM35 User Manual
Page 20
3.8 Battery
replacement
• Open the battery compartment and insert 3 x 1,5 V Mignon cell AA, IEC LR6.
Always replace all three batteries at the same time!
Ensure correct polarity!
Dispose of used batteries in accordance with environmental
regulations
!
3.9 Calibration
Re-calibration may only be carried out by specialist staff and with the corresponding
pressure standards.
We recommend that you have the HM35 re-calibrated at least once a year, and, in case
of highest demands for precision, several times a year.
Manual re-calibration
• In
the
Menu,
select the Calibration
→ Manual Calibration function.
Zero point
(Offset)
1
. Open the pressure connection or, with the absolute pressure unit, set the given
pressure value to the normal pressure.
2.
Press
the
Function
key (
)
Î
the zero point will be re-calibrated.
Full-scale value
1
.
Set the given pressure value to the normal pressure.
2
. Press
the
Function
key (
)
Î
the full-scale value will be calibrated and the HM35 returns to normal operation.
19