2 ir-hardware, Of the instrument – Dwyer HM35 User Manual

Page 13

background image

12

Return Value

Command processed:
Return Value = o k
Example for response:

o k HT * 1 3 CR


Error

Return Value

Meaning

er-001

RS232 Protocol checksum Error

er-110

Header Error; Too short
Header Error; Too many subnodes
Header Error; Query not at leaf node
Header Error; Multiple querys
Header Error; Characters after query
Header Error; Too long

er-113

Undefined Header; Undefined command

er-109 Missing

parameter

Missing parameter; Boolean expected
Missing parameter; String expected
Missing parameter; Discrete expected
Missing parameter; Not of expected type

er-101

Invalid character; Terminator expected

er-108

Invalid parameter; Out of bounds
Invalid parameter; Too long

er-203 Command

Protected

er-999

EEProm Read/Write Error

er-002

Fatal Command Execution Error

Example: Checksum

Error

Response:

e r – 0 0 1 HT * 200 CR


After command with response value

Example: reading time (07:08:09)

Response:

0 7 , 0 8 , 0 9 HT * 1 9 5 CR


After every command wait for the response of the instrument (max. 680 ms).


3.7.2

IR-Hardware of the instrument


The hardware of the IR-connection of the instrument is compatible with IrDA- Standard
1.0

.


IR (IrDA)- Adapter

A passive IrDA- adapter has to be used which is compatible to IrDA- Standard 1.0. The
IrDA adapter ACT-IR220Lplus is available as accessory.
The following explanations apply to this type: