beautypg.com

Dwyer HM35 User Manual

Page 18

background image

17

Parameter
limits

Feedback data

Feedback data
type

Description

Remarks

3min, 10min, 60min ---

---

Set auto Off time

---

---

---

Switch off instrument

ON, OFF

---

---

Enable/disable auto zero

25M./s, 10M./s,
1s ... 24h

---

---

Set interval time for Data
Logging

Similar to Menu

10s, 30s ... 24h

---

---

Set time period for
Average

Similar to Menu

OFF

---

---

Set auto IrDA connection

---

,,... INTEGER

Query the BIT-Error
memory

Variable amount of
Feedback data, err: 0...255

DISCRETE

Query the BIT-error
memory

If Error memory deleted:
'No Errors!’

OFF, level 1,
level 2, level 3

---

---

Brightness setting for
LCD backlight

EX-instrument only OFF
and Level 1, similar to
Menu

---

diverse

Select Memory Data
Logging (cyclic)

Complete reading,
Format: see 3.6.1

---

---

---

Delete Memory Data
Logging

---

---

---

Delete Status and Error
Memory

---

STRING Query

instrument

identification

See SYST:IDENT?

---

INTEGER

Query Status Byte

---

INTEGER

Initiate a Self-Test

---

---

---

Reset command