beautypg.com

Dwyer 629 User Manual

Series 629 differential pressure transmitter, Dwyer instruments, inc

background image

DWYER INSTRUMENTS, INC.

Phone: 219/879-8000

www.dwyer-inst.com

P.O. BOx 373 • MIChIgAN CITY, INDIANA 46360, U.S.A.

Fax: 219/872-9057

e-mail: [email protected]

Series 629 Differential Pressure Transmitter

Specifications - Installation and Operating Instructions

Bulletin E-112

1/4 NPT

TYP

6-59/64 [175.82]

OPTIONAL 1/4 NPT

MALE FITTING

5-25/32 [146.84]

1/4 NPT FEMALE FITTING

1-5/8

[41.28]

3-9/32

[83.34]

1-11/32

[34.13]

43/64

[17.07]

LIQUID TIGHT FITTING

CORD DIAMETER RANGE:

.200 TO .350 [5.06 TO 8.89]

Range
(psid)
5
10
25
30
50
100
150
200
300
500

Working*
Pressure (psi)

10
20
50
60
100
200
300
400
600
1000

Over
Pressure (psi)

50
100
250
250
250
500
750
1000
1200
2000

* Pressures exceeding the working
pressure limit may cause a calibration
shift of up to ±3% of full scale.

Pressure Limits

The Series 629 Differential Pressure Transmitter monitors differential
pressure of air and compatible gases and liquids with 0.5% accuracy.
The design employs dual pressure sensors converting pressure
changes into a standard 4-20 mA output signal for two wire circuits.
Small internal volume and minimum moving parts result in exceptional
response and reliability. Terminal block, zero and span adjustments are
easily accessed under the top cover. The Series 629 Differential
Pressure Transmitter is designed to meet NEMA-4X (IP66) construction.

SPECIFICATIONS
Service:
Compatible gases & liquids.
Wetted Materials: Types 316, 316L SS.
Accuracy: ±0.5% F.S. (includes linearity, hysteresis & repeatability).
Temperature Limits: 0 to 200°F (-18 to 93°C).
Compensated Temperature Limits: 0 to 175°F (-18 to 79°C).
Pressure Limits: See chart.
Thermal Effect: 0.02%/°F (0.036%/°C) includes zero & span.
Power Requirements: 13-30 VDC.
Output Signal: 4 to 20 mA. Optional 0-5, 1-5, 0-10, 1-6 or 2-10 VDC.
Zero & Span: Potentiometers inside conduit enclosure.
Response Time: 50 msec.
Loop Resistance: 0-1300 ohms maximum for current output. For
voltage outputs, minimum load resistance: 2000 ohms.
Current Consumption: Transmitter: 40 mA max; Transmitter with
optional display: 140 mA max.
Electrical Connections: Terminal block.
Process Connections: 1/4˝ female NPT.
Display: Optional 4-1/2 digit 1/2˝ (12.7 mm) height LED.
Enclosure Rating: Designed to meet NEMA 4X (IP66) for non-LED
models.
Mounting Orientation: Not position sensitive.
Weight: 10.1 oz (286 g).
Agency Approvals: CE:
EN 61000-4-2
EN 61000-4-3
EN 61000-4-4
EN 61000-4-5
EN 61000-4-6

Max. Static on Either Port