beautypg.com

Operation and maintenance, Generating pressure – Crystal T-1-CPF Hydraulic Comparator User Manual

Page 10

background image

Operation and Maintenance

8

Type T Pump Operation Manual

Operation and Maintenance

„„„„

„

GENERATING PRESSURE

Note:

It may be necessary to prime the pump, particularly if the pump has been inactive for an extended period of time. Refer to

Priming on page 8

for instructions on priming the pump.

1

Check that the vent fill plug is fully open by turning it counter-clockwise.

2

Open the vent valve.

Open the vent plug…

then open the vent valve.

3

Set the valve rod selector to control your desired fluid output volume.

!

CAUTION: Only move the valve rod selector while the pump handle is stationary, in the up position.

Note:

With the valve rod in the “out” position, fluid flow will be at maximum and the maximum achievable pressure will be 1000 psi. To exceed 1000 psi, push

the valve rod in.

Low Pressure/

High Volume

High Pressure/

Low Volume

This manual is related to the following products: