Differential Pressure Measurement Uncertainties without Tare

The total nVision Reference Calibrator measurement uncertainty in the ΔP mode configuration will need to consider the uncertainties of both pressure mod-
ules. We recommend the module uncertainties to be combined with the preferred square root of the sum of the squares (or “root sum squares”) method.

The following table lists the possible combinations of combining Pressure Modules (PM) with different accuracy statements. The uncertainties reported below
are without using the Tare feature which will greatly improve your measurement uncertainty.

Upper Pressure Module Uncertainties
(of Static Line Pressure) (of Reading)

0.025%

0.05%

0.10%

Lower Pressure
Module Uncertainties
(of Static Line Pressure)
(of Reading)

0.025%

0.035%

0.056%

0.103%

0.05%

0.056%

0.071%

0.112%

0.10%

0.103%

0.112%

0.141%

Differential Pressure Measurement Uncertainties with Tare

The Tare function can improve measurement uncertainties on two modules with the same full scale pressure range installed into one nVision Reference
Recorder.

The following specifications apply to the measurement system with a logging interval of 1 reading/second or slower:

Full Scale Range of Both Sensors

The Greater of (+/–)

PSI

bar

kPa/MPa

kg/cm

2

PSI

mbar

inH

2

O

mmH

2

O

% of DP Reading

30

3

300

3

0.0005

0.04

0.014

0.4

or

0.025%

100

10

1

10

0.0015

0.10

0.04

1.0

or

0.025%

300

30

3

30

0.005

0.4

0.14

4.0

or

0.025%

1000

100

10

100

0.02

1.0

0.4

10.0

or

0.05%

3000

300

30

300

0.05

4.0

1.4

n/a

or

0.05%

10000

700

70

700

0.2

10.0

4.0

n/a

or

0.1%

15000

1000

100

1000

0.3

15.0

6.0

n/a

or

0.1%

Unit must be enabled in CrystalControl