Crystal 30 Series Digital Pressure Calibrator User Manual
Page 9

Specifications
7
30 Series Operation Manual
Power
Battery � � � � � � � � � � � � � � � � � � � � � � � � � � One Alkaline or Lithium 9V battery
X
Approved Batteries
ATEX/IECEx:
The 30 Series is Intrinsically Safe only if powered by one of the following battery types:
Approved Battery Type
Ta=
Marking
Duracell Alkaline 9V, MN 1604
0 to 50° C
Ex ia IIC T4 Gb
Energizer Alkaline 9V, 522
Energizer Lithium 9V, L522
Varta High Energy Alkaline 9V, 4922
0 to 45° C
Many other battery types and models have been tested but failed to meet the requirements for potentially explosive atmospheres—do not assume other
models are equivalent�
Do not mix battery types or manufacturers�
CSA:
Substitution of components may impair intrinsic safety�
CSA:
La substitution de composants peut compromettre la sécurité intrinsèque�
CSA:
The 30 Series is Intrinsically Safe only if powered by one of the following battery types:
CSA:
Le 30 Series est un système à sécurité Intrinsèque seulement s’il est alimenté par un des Piles de type suivant:
Approved Battery Type
(Approuvé Type de Batterie)
Marking
(Marquage)
Duracell Alkaline 9V, MN 1604
Class I, Division 1, A, B, C, D T4
(Classe I, Division 1, A, B, C, D T4)
Energizer Alkaline 9V, 522
Energizer Lithium 9V, L522
Varta High Energy Alkaline 9V, 4922
Class I, Division 1, A, B, C, D T3C
(Classe I, Division 1, A, B, C, D T3C)
Battery Life � � � � � � � � � � � � � � � � � � � � � � 90 hours typical
Low Battery Indicator � � � � � � � � � � � � Flashing battery icon
Connections
Pressure Fitting � � � � � � � � � � � � � � � � � � 1/8" female NPT
Milliamperes � � � � � � � � � � � � � � � � � � � � 4mm jacks, 19mm (0�75”) spacing
Communication � � � � � � � � � � � � � � � � � micro USB
!
WARNING: Do not use the micro USB serial interface in hazardous locations.