Crystal 30 Series Digital Pressure Calibrator User Manual
Page 8

Specifications
6
30 Series Operation Manual
Media Compatibility (LP and HP)
Liquids and gases compatible with PTFE (Polytetrafluoroethylene) penetrated, hard anodized aluminum (transducer housing), 316 stainless steel (sensor), and
Viton
®
(O-ring). (Viton is a registered trademark of Dupont Dow Elastomers.)
Products with the letters “SPSS” in the suffix of the part number (e.g., “IS33-36/3000-SPSS”), have stainless steel manifolds and therefore the only wetted mate-
rials are stainless steel and Viton.
Pressure Conversions
1 PSI =
27.6806 inches of water column (water at 4°C [39.2°F])
703.087 millimeters of water column (water at 4°C [39.2°F])
70.3087 centimeters of water column (water at 4°C [39.2°F])
2.03602 inches of mercury (mercury at 0°C [32°F])
51.7149 millimeters of mercury (mercury at 0°C [32°F])
6.8948 kilopascals
0.070307 kilograms per square centimeter
0.068948 bar
68.948 millibar
0.0068948 megapascals
Note:
Other conversions may have been specified at time of order. Refer to your certificate of calibration for details.
Electrical
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 55 mA
Resolution . . . . . . . . . . . . . . . . . . . . . . . 0.001 mA
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . ±(0.025% of reading + 0.001 mA)
Maximum Voltage . . . . . . . . . . . . . . . 30 VDC
Maximum Current . . . . . . . . . . . . . . . 100 mA
Note:
When connecting to circuits that included HART transmitters, a 250 ohm resistor must be placed in series with the calibrator, to prevent damage to the
calibrator and any other equipment connected to the loop. mA can be displayed as a percentage, where 0 to 100% corresponds to either 4 to 20 mA or
10 to 50 mA.