Safety & certifications, Hazardous locations, Certifications – Crystal 30 Series Digital Pressure Calibrator User Manual
Page 11

Safety & Certifications
9
30 Series Operation Manual
Safety & Certifications
HAZARDOUS LOCATIONS
Every 30 Series pressure calibrator includes the following Intrinsic Safety approvals.
II 2G Ex ia IIC T4 Gb
FTZU 06 ATEX 0010X
This product conforms to the following standards:
EN60079-0: 2006 | EN60079-11: 2007
Ex ia IIC T4 Gb
IECEx FTZU 10.0018X
This product conforms to the following standards:
IEC 60079-0: 2007 | IEC 60079-11: 2006
Exia Intrinsically Safe and Non-Incendive for Hazardous
Locations: Class I, Division 1, Groups A, B, C and D,
Temperature Code T4/T3C.
Sécurité intrinsèque et non incendiaire pour dan-
gereux Lieux: Classe I, Division 1, Groupes A, B, C et D,
Code de température T4/T3C.
Milliamp Input Entity Parameters:
Ui = 30V
Uo = 9.9V
Entity parameters specify the safe voltage, current,
capacitance and inductance that can either be
connected to the device, or is internal to the device.
Ta = 0 to 50° C
Ii = 100mA
Io = 2.62mA
Pi = 0.75W
Po = 6.5mW
Ci = 0
Co = 3.2uF
Li = 0
Lo = 100uH
!
WARNINGS: The following warnings apply to the 30 Series Pressure Calibrator:
•
Substitution of components may impair intrinsic safety.
•
Rep
lace batteries with approved type in non-hazardous locations only.
•
Do not use USB interface in hazardous locations.
CERTIFICATIONS
The 30 Series has been tested and certified to comply with a variety of international standards.
This 30 Series complies with the Australian requirements
for the C-tick mark. The instrument was tested against
AS/NZS 3584, C-tick EMC/EMI requirements.
Crystal Engineering declares that the 30 Series is in ac-
cordance with the ATEX Directive, the Electromagnetic
Compatibility Directive, and the Pressure Equipment
Directive per our declaration(s).
This 30 Series is approved for use as a portable test
instrument for Marine use and complies with Det
Norsjke Veritas’ Rules for Classification of Ships, High
Speed & Light Craft and Offshore Standards.
C-tick
ATEX