Badger Meter M-Series M1000 User Manual
Installation manual, M-series® m1000
M-Series® M1000
Data Logging
MAG-UM-00677-EN-02 (July 2014)
Installation Manual
DESCRIPTION
The Data Logging kit can be ordered with this part number:
67354-010
The Data Logging feature records three types of events:
•
Totalizer/error events
•
Configuration change events
•
Startup events (power up or reset events)
Each type of event is recorded into three separate files stored on
internal memory.
OTEE:
N
Over time the data logging will reach the capacity of the
memory. Any new events to be recorded will overwrite
the oldest event on record.
Totalizer/Error Events
The capacity of the logging memory is 30,000 messages. The table
below defines the capacity of the memory configured for data
logging. On each interval the totalizers are recorded in addition to
any errors that have occurred from the last interval. To program the
interval, go to Miscellaneous > Datalog Period.
Interval
Totalizer / Error Events
15 min
up to 312 days
1 hr
up to 1250 days
6 hr
up to 20 years
12 hr
up to 40 years
24 hr
up to 80 years
Configuration Events
Each Configuration event identifies the parameter that was
modified and to what value the parameter was changed. Up to 40
configuration events can be recorded.
Startup Events
Each Startup event identifies the time and reason of the event. The
M1000 does not record the date and time of a power off. A total of
20 startup events can be recorded.
EXTRACTING THE EVENT FILES
All logged events can be extracted from the meter using the
supplied Flow Meter Tool software, which connects the meter to a
laptop via the supplied RS232 cable.
RS232 Link File Extraction
1. Identify / Configure the meter’s communication settings:
a. Navigate to Communications > Interface.
b. Set the interface to MODBUS.
OTEE:
N
The interface must be set to MODBUS. All other settings
can be set as desired by the operator and must match
those settings of the software tool.
c. Record or change other interface parameters
(parity and baud rate).
2. Connect the supplied RS232 cable into the RS232 connector of
the meter. Either connect the serial connector to a COM port or
connect it to the USB adapter.
3. Open the Flow Meter Tool installed on the laptop or PC. Go to
Start > All Programs > Badger Meter to open the Flow Meter Tool
application.
4. To configure the Flow Meter Tool software communication
settings:
a. Select Communication Settings.