Page 39: Physical

background image

Operating Conditions

Ambient Temperature:

0

°

C to 55

°

C

Relative Humidity:

20% - 95% non-condensing

Performance Under Operating Conditions

Temperature Stability:

0.005% of span/

°

C change in ambient

temperature.

A.10

ENVIRONMENTAL

EMI Susceptibility:

Complies with EN50082-1:1992,

EN50082-1:1995.

NOTES:

1. For RF electromagnetic fields (10V/m 80% AM 1kHz), the reading
accuracy may be impaired by up to

0.3% in the frequency band

87 to 109MHz.

2. For line-conducted disturbances induced by RF fields (10V 80%

AM 1kHz), the product is self-recoverable in the frequency band

0.15 - 0.73MHz.

EMI Emmissions:

Complies with EN50081-1:1992,

EN50081-2:1993.

Safety:

Complies with EN61010-1.1993.

Supply Voltage:

90 - 264V, 50/60Hz (standard)

20 - 50V AC/DC (option)

Power Consumption:

4watts approximately

Front Panel Sealing:

To IP66 (Nema 4)

A.11

PHYSICAL

Dimensions:

Height - 48mm

Width - 96mm

Depth - 100mm

Mounting:

Panel-mount; press-fit fixing strap supplied.

Panel cut-out - 45mm x 92mm

Terminals:

Screw type; combination head.

Weight:

0.21kg maximum.

PM091-A

Volume II

A-4

59136