beautypg.com

Thermo Fisher Scientific Ion Selective Electrodes Copper User Manual

Page 8

background image

Instruction Manual

Copper Electrode

8

convenient concentration unit. Table 1 indicates some of these
concentration units.

TABLE 1

: Concentration Unit Conversion Factors


ppm

Cu

+2

moles/liter


6354.0 1.0X10

-1


635.0

1.0X10

-2


63.5

1.0X10

-3


6.4

1.0X10

-4


MEASUREMENT PROCEDURES

Direct Measurement

A simple procedure for measuring a large number of samples. A
single meter reading is all that is required for each sample.
The ionic strength of samples and standards should be made the
same by adjustment with ISA. The temperature of both sample
solutions and standard solutions should be the same.

Direct Measurement of Cupric Ion (using a standard pH/mV meter)

1. By serial dilution of the 0.1M or 1000 ppm standards,

prepare 10

-2

, 10

-3

, and 10

-4

M or 100 and 10 ppm standards

for the cupric ion. To 100 ml of each standard, add 2
ml of ISA. Prepare standards with a composition
similar to the samples if the samples have an ionic
strength above 0.1M.


2. Place 100 ml of the 10

-4

M or 10 ppm standard in a 150 ml

beaker, place the beaker on the magnetic stirrer and
begin stirring at a constant rate.


3. Lower the electrode tip(s) into the solution. Make

sure that the meter is in the mV mode. When the
reading has stabilized, record the mV reading.


4. Place 100 ml of the 10

-3

M or 100 ppm standard in a 150

ml beaker, place the beaker on the magnetic stirrer and
begin stirring at a constant rate.


5. After rinsing the electrode(s) with distilled water and

blotting dry, immerse the electrode tip(s) in the
solution. When the reading has stabilized, record the
mV reading.