Thermo Fisher Scientific Ion Selective Electrodes Copper User Manual
Page 24

24
in solution
in solution
Noisy or Unstable
defective meter
check meter with
Readings (readings
shorting strap
continuously or
rapidly changing)
air bubble on
remove bubble by
membrane
redipping
electrode
electrode exposed
soak electrode in
to interferences
cupric standard
defective electrode
replace electrode
ISA not used
use recommended ISA
meter or stirrer
ground meter or
not
grounded
stirrer
Drift (reading
samples and standards allow solutions to
slowly changing
at different
come to room temper-
in one direction)
temperatures
ature before
measurement
electrode exposed
check section
to
complexing
entitled
agents
Precipitation and
Complexation
incorrect reference
use recommended
filling solution
filling solution
Low Slope or
standards contamin-
prepare fresh
No Slope
ated or incorrectly
standards
made
ISA not used
use recommended
ISA
standard used as ISA
use ISA
electrode exposed
check section
to complexing agents
entitled
Precipitation
and Complexation
air bubble on
remove bubble by
membrane
redipping
probe
"Incorrect Answer"
incorrect scaling
plot millivolts on
(but calibration
of semi-log paper
the linear axis. On
curve is good)
the log axis, be