Noisy or Unstable

defective meter

check meter with

Readings (readings

shorting strap

continuously or
rapidly changing)

air bubble on

remove bubble by

membrane

redipping

electrode


electrode exposed

soak electrode in

to interferences

cupric standard


defective electrode

replace electrode


ISA not used

use recommended ISA


meter or stirrer

ground meter or

not

grounded

stirrer


Drift (reading

samples and standards allow solutions to

slowly changing

at different

come to room temper-

in one direction)

temperatures

ature before

measurement

electrode exposed

check section

to

complexing

entitled

agents

Precipitation and

Complexation

incorrect reference

use recommended

filling solution

filling solution


Low Slope or

standards contamin-

prepare fresh

No Slope

ated or incorrectly

standards

made

ISA not used

use recommended

ISA

standard used as ISA

use ISA


electrode exposed

check section

to complexing agents

entitled

Precipitation

and Complexation


air bubble on

remove bubble by

membrane

redipping

probe


"Incorrect Answer"

incorrect scaling

plot millivolts on

(but calibration

of semi-log paper

the linear axis. On

curve is good)

the log axis, be