Standardize – Thermo Fisher Scientific CyberScan DO 1500 User Manual

Page 41

background image

Instruction Manual

CyberScan DO 1500

38

5.4.

mg/L– One Point calibration (Auto)


User can do calibration in mg/L.

1. Press

std to enter to ‘Standardize’ screen. Upper display shows

the present measured value while lower display shows theoretical
value. [Based on pressure and salinity setting]

2. User can just confirm the theoretical reading.

3. Acceptance window is 70% of the theoretical value.

4. Press

enter key to confirm calibration. Meter blinks the cal values

for few seconds before exiting to measurement mode.

Standardize

STABLE

mg/L

press

to

accept

enter