Instruction Manual

CyberScan DO 1500

37

5.3.

%Sat – One Point calibration (Manual)


User can calibrate to the know value.

Accepted window is 70% of the default value. Lowest value that can
be calibrated is 20.0 %Sat.

1. Press

std to enter to ‘Standardize’ screen. Both Upper and lower

display shows the present measured value.


2. Dip the electrode into the standard solution. Wait for the upper

display value is stabilised. Use /

keys to adjust the lower

display to the known cal value.


3. Press

enter to confirm the reading. Meter blinks the cal values for

few seconds before exit to measurement mode.

Standardize

% Sat

press

to

accept

enter

press

to set value

STABLE